Nuclear Technology / Volume 168 / Number 3 / December 2009 / Pages 591-595
Nuclear Data / Special Issue on the 11th International Conference on Radiation Shielding and the 15th Topical Meeting of the Radiation Protection and Shielding Division (PART 3) / Radiation Measurements and Instrumentation / dx.doi.org/10.13182/NT09-A9274
Articles are hosted by Taylor and Francis Online.
For the International Thermonuclear Experimental Reactor (ITER), the European Union intends to introduce a test blanket module using liquid lithium lead. In the present study, a preliminary experiment was conducted using a LiAl/Pb assembly, which has a neutronics performance similar to lithium lead. The neutron spectrum at the inside of the assembly was measured with an NE-213 detector. We found that the spectra calculated with MCNP5 and JENDL-3.3 underestimate the 14-MeV peak by no less than 30%, while ENDF/B-VII.0 and JEFF-3.1 gave good agreement. The neutron nuclear data of lead stored in JENDL-3.3 may have some problems.