American Nuclear Society
Home

Home / Publications / Journals / Nuclear Science and Engineering / Volume 13 / Number 4

An Improved Procedure for U235 Isotopic Ratio Analysis in Impure Materials

W. D. Kelley, B. L. Twitty

Nuclear Science and Engineering / Volume 13 / Number 4 / August 1962 / Pages 374-377

Technical Paper / dx.doi.org/10.13182/NSE62-A26179

An improved neutron activation procedure for determining the U235 content in impure uranium samples is presented. With 95% confidence, a relative precision of ±1% is obtained at the level of 0.7 wt % U235. An ethyl acetate extraction is used to purify the uranium chemically. Most of the uranium decay products are removed by extraction with 2-thenoyltrifluoroacetone (TTA) in xylene. The purified uranium, which is subsequently activated as U3O8, has a consistent gamma background level. Utilization of a i.d. well scintillation crystal has permitted the use of large diameter aluminum planchettes holding 2 gm aliquots of U3O8. This system has minimized the problem of varying irradiation geometries caused by differences in U3O8 bulk density. Data from a sample recycle program are presented to illustrate the relative precision of ±1%. Comparisons between neutron activation and mass spectrometric results agree within ±1%.