American Nuclear Society
Home

Home / Publications / Journals / Fusion Science and Technology / Volume 48 / Number 1

A Compact Diffusion Sampler for Environmental Applications Requiring HTO or HTO + HT Determinations

R. L. Otlet, A. J. Walker, I. D. Mather

Fusion Science and Technology / Volume 48 / Number 1 / July-August 2005 / Pages 366-369

Technical Paper / Tritium Science and Technology - Tritium Measurement, Monitoring, and Accountancy / dx.doi.org/10.13182/FST05-A944

Passive diffusion samplers have now been in use in the UK for the measurement of tritium at environmental levels in the form of HTO and HTO + HT for more than ten years. At the outset their main purpose was to identify the direction of a possible inadvertent release of tritium into the environment. More recently, however, there has been growing interest in their use as stand-alone devices for tritium determination at environmental levels. This has necessitated a more rigorous design to reduce the "draughts effect" in exposed locations and has also required modifications to enable the measurement of HTO + HT.

The paper describes the recent design improvements and the results of validation tests carried out against dynamic, discriminating (dry-bed) samplers.