American Nuclear Society
Home

Home / Publications / Journals / Fusion Science and Technology / Volume 10 / Number 3P2A

Techniques for Measurement of Velocity in Liquid-Metal MHD Flows

C. B. Reed, B. F. Picologlou, P. V. Dauzvardis, J. L. Bailey

Fusion Science and Technology / Volume 10 / Number 3P2A / November 1986 / Pages 813-821

Liquid-Metal Blankets and Magnetohydrodynamic Effects / Proceedings of the Seveth Topical Meeting on the Technology of Fusion Energy (Reno, Nevada, June 15–19, 1986) / dx.doi.org/10.13182/FST86-A24839

Three instruments for measuring local velocities in liquid-metal MHD experiments for fusion blanket applications are being evaluated. The devices are used in room-temperature NaK experiments to measure three-dimensional flow field patterns anticipated in complex blanket geometries. Hot film anemometry, a standard technique in ordinary fluids, is being used, as well as two developmental devices. One is called the Liquid Metal Electromagnetic Velocity Instrument (LEVI), and performs essentially as a local DC electromagnetic flow meter. The third device, a Thermal Transient Anemometer (TTA) is a rugged, yet relatively simple device, which measures local velocity through the mechanism of convective heat transfer, in some ways similar to hot-film anemometry. Results are presented showing the kinds of data collected thus far with each instrument. Measurements include both local velocity measurements and some preliminary frequency analyses of the fluctuating signals from both a hot-film sensor and the LEVI device.