A New Tritium Monitor for the Tokamak Fusion Test Reactor

R. A. Jalbert

Fusion Science and Technology / Volume 8 / Number 2P2 / September 1985 / Pages 2077-2081

Monitoring and Measurement / Proceedings of the Second National Topical Meeting on Tritium Technology in Fission, Fusion and Isotopic Applications (Dayton, Ohio, April 30 to May 2, 1985) / dx.doi.org/10.13182/FST85-A24590

At DT-fueled fusion reactors, there will be a need for tritium monitors that can simultaneously measure in real time the concentrations of HTO, HT and the activated air produced by fusion neutrons. Such a monitor has been developed, tested and delivered to the Princeton Plasma Physics Laboratory for use at the Tokamak Fusion Test Reactor (TFTR). It uses semipermeable membranes to achieve the removal of HTO from the sampled air for monitoring and a catalyst to convert the HT to HTO, also for removal and monitoring. The remaining air, devoid of tritium, is routed to a third detector for monitoring the activated air. Ihe sensitivities are those that would be expected fron tritium instruments employing conventional flow-through Ionization chambers: 1–3 µCi/m3. Its discriminating ability is approximately 10−3 for any of the three components (HTO, HT and activated air) in any of the other two channels. For instance, the concentration of HT in the HTO channel is 10−3 times its original concentration in the sampled air. This will meet the needs of TFTR.