American Nuclear Society
Home

Home / Publications / Journals / Fusion Science and Technology / Volume 59 / Number 1

Metrology Statistics for NIF Tuning Campaign

H. Huang, R. Fallon, S. W. Haan, Y. T. Lee, K. A. Moreno, A. Q. L. Nguyen, A. Nikroo, K. L. Sequoia, R. B. Stephens, J. J. Wu

Fusion Science and Technology / Volume 59 / Number 1 / January 2011 / Pages 26-34

Technical Paper / Nineteenth Target Fabrication Meeting / dx.doi.org/10.13182/FST10-3694

Delivering targets whose parameters are repeatable within specifications is critical to a successful ignition campaign. Knowing the repeatability of the metrology techniques and determining random error bars is central in achieving this goal. In this paper, we will discuss the relationship of specifications and actual target variability to the processes we have developed to adequately characterize them. The paper covers five broad categories of capsule specifications: dimension, composition, homogeneity, surface finish, and point defects. The instruments include contact radiography, energy dispersive spectroscopy, X-ray absorption edge spectroscopy, precision radiography system, sphere mapping, and phase-shifting diffraction interferometry.